MPC-9000_Series

 

  • Manual Single Sample Counter

  • 5.1cm (2 in.) Standard Planchet Diameter

  • Dual Phosphor Scintillator Detector

  • Touch Screen

  • Light Weight Portability

  • Radon Compensation For Air Filters

  • Other MPC-900 Models:

 

Overview

The MPC-900-DP is a convenient and versatile “benchtop” counter equipped with a dual phosphor scintillator type detector that supports gasless operation and easy portability. This sample counter is ideal for standard Health Physics applications such as smears and air filters.

 

Specifications

Specifications are subject to change without notice

TYPE    
  Sample Changer Manual
  Number of Detectors 1
  Counter Gross
  Sample Size 5.1 cm (2 in.) Diameter
  Interface Touch Screen
DETECTOR  
  Size 5.1 cm (2 in.) Diameter
  Type ZnS+plastic dual phosphor
  Window Aluminized, 80 µg/cm2
  Shielding (Optional)
     
SAMPLE  
  Size 5.1 cm (2 in.) Diameter planchets
  Capacity 1
  Depth 1/8, 1/4, 5/16 inch
  Data Export USB Flash Drive, RS-232
     
INTERFACE  
  Type Touch Screen
  Compatible Software DUO Software
     
PERFORMANCE  
    TYPICAL        WARRANTED
  Alpha Background 0.2 cpm          0.3 cpm
  Beta Background 40 cpm           65 cpm
  Th230 Efficiency 23%
  Sr90/Y90 Efficiency 45%
  Tc99 Efficiency 30%
     
  Alpha-Beta Crosstalk < 0.1%
  Beta-Alpha Crosstalk < 0.1%
  Count Modes Alpha Only, Beta Only, Alpha/Beta Simultaneously
     
PHYSICAL  
  Size 45.7 x 22.9 x 40.6 cm (18 x 9 x 16 in.) (H x W x D)
  Weight 18.1 kg (40 lb) Installed

 

Compatible Protean Application Software Packages

  • DUO Software

 

Options

  • Lead ring shielding, 22.7 kg (50 lb), Order part number MPC-900-SHLD
  • Remote PC Interface (without touch screen) Order Model MPC-900-PC-GFW-2

 

Downloads